HOME > Products > sunsepTM > Technical Support

Technical Support:
sunsepTM Membrane Type Gas Dryer

メンブレン式ドライヤー サンセップ(sunsep)

AGC ENGINEERING CO.,LTD.

Technical Support

When specifications and on OEM contact are required, please feel free to consult with us.

North America

BELLEX INTERNATIONAL CORPORATION
200 Bellevue Parkway, Suite 180 Wilmington, DE 19809
Phone : +1-302-791-5180
Facsimile : +1-302-791-5190

China

SHANGHAI TSINGROSS ENVIRONMENTAL TECHNOLOGY CO.,LTD.
2-23, No.5808 Shenzhuan Rd. Songjiang District, Shanghai. P.C.201619
Phone : +86-21-61196289-12
Facsimile : +86-21-61196290-10
E-Mail : louyouyi@tsingross.com

Japan and others

AGC Engineering Co., Ltd.
Membrane Div.
Customer Service Group.
WBG MARIVE WEST 19F, 2-6-1 Nakase, Mihama-ku,
Chiba-shi, CHIBA 261-7119, JAPAN
Phone : +81-43-350-3378
Facsimile : +81-43-350-3383
E-mail : sunsep.agc@agc.com

Catalog

=>Download PDF File

Instructions for Use

  • To remove drainage, dust , etc. , we recommend that a filter (filtering capacity of less than 5µm) be installed at the inlet line of supply gas. Dehumidifying efficiency may decrease if drainage is mixed into the inlet line. We therefore recommend that an automatic drain system be installed at the supply gas inlet.
  • When a sunsepTM unit is used for the line from a lubricating type compressor, we recommend that an automatic oil mist separator be installed at the inlet line of supply gas. The oil mist-separator must have an oil filterability capability of less than 0.3µm ( 95% collection capability of dust diameter ) and should also have an oil-mist concentration of less than 1 mgf/Nm3.
  • Supply gas should be clean and free of dust, corrosive gas, organic solvent, and chemicals.
  • Excessive pressure should not be applied to the purging gas inlet and outlet.(max. 7.1psig)
  • For operation with a small amount of purging gas, not more than 0.35 scfm. we recommend that a small orifice be used instead of a needle valve.
  • When installing a pressure reducing valve, we recommend that the valve be installed at the outlet line of dry gas, which will increase dehumidifying efficiency.